Configuring the Mannequin
25
Installing the Support Tub
The support tub is used to hold the fetus during prepartum/latent phase.
1. Install the Leopold support tub in the mannequin’s torso.
The Mannequin’s Torso - The Support Tub
The mannequin is ready for installing the Leopold fetus.
CAUTION:
Do not place pressure or any object on the rotation ring support
arms while the rotation ring is removed.
Loading the Fetus - Prepartum
Prepartum and Latent
To load the Leopold fetus:
2. Ensure the Leopold support tub is installed in the mannequin’s torso.
3. Place empty placenta pouch on top of the support tub.
4. Place the Leopold fetus on top of the placenta pouch and tub in the
desired position.
5. Cover with the prepartum abdomen.
Active Phase
To load the delivery fetus:
1. Install the placenta pouch.
2. Attach umbilical cord to fetus and placenta. Place placenta inside pouch.
3. Place the delivery fetus on top of the placenta pouch.
4. Cover with the prepartum abdomen.
The
rotation
ring removed