Wireless Voice Link
341
Preparing the Base Station in the Simulator
When using the base station in the simulator, ensure the batteries are removed and the following items
are attached:
• Power cable
• Line out cable
The DIP switch is located in the battery compartment of the base station (Figure 5: DIP Switch Settings).
Dip Switch Settings
The base station should come already connected and installed inside the simulator.
To prepare the base station:
1. Set the base station DIP switch positions 6 and 7 to OFF, and 8 to ON.
2. Turn the power off and on using the power switch on the outside of the
base station to ensure the DIP switch changes take effect.
3. Leave the power switch on the outside of the base station in the on
position.
DIP Switch Settings for the Base Station
NOTE:
Since the base station receives power from the simulator, the power switch on the outside of the
base station must remain in the ON position. Use this power switch to refresh DIP switch settings. Do
not turn the simulator off and on to refresh the DIP switch settings.
DIP switch settings