4.6 Jitter Tolerance Test
4-29
4
O
per
ation
4.6 Jitter Tolerance Test
4.6.1 Jitter Tolerance Test Setup Screen
On the Run Test tab, you can view the following Jitter Tolerance settings.
The references for each setup area are shown in the following figure.
Figure 4.6.1-1 Jitter Tolerance Setup Screen
Table 4.6.1-1 Measurement Sequence Setting Items
Item
Description
Measurement
Sequence
Specifies the measurement sequence direction.
From higher Freq.side: Measure s from the higher modulation
frequency side.
From lower Freq.side: Measures from the lower modulation
frequency side.
BitRate
Displays the current bit rate.
Clock Selection* Selects Recovered Clock or External Clock for the clock.
*: Can be changed when started in PCIe Link Sequence and the
MU18304xB with Option 22 or 23 installed is selected for
Equipment.
Not displayed when started by Jitter Tolerance Test.
Bit Rate
Clock Selection
Menu
Table 4.6.1-2 Measurement Status
4.6.5 Jitter Tolerance Measurement Start
Measurement Sequence