Chapter 3 Connecting Equipment
3-14
1. When MX183000A is installed on a control PC, connect the control
PC to MP1800A with an Ethernet cable. Or connect the MP1800A to
the MP1825B with a USB cable.
When MX183000A is installed on MP1800A, Ethernet cable
connection is not required.
MP1800A requires the MP1800A-x02 LAN option.
2. Set up as follows using the Remote Control tab on Setup Utility.
Activate Interface:
Ethernet
Performance:
Enhanced
3. Mount MU181000A/B-x01 and MU181500B in MP1800A-x02, x32.
4. Mount MU183020A-x30 in Slot 3 of MP1800A-x02, x32.
5. Connect the Clock Output connector of MU181000A/B to the Ext
Clock Input connector of MU181500B with a coaxial cable.
6. Connect the Jittered Clock Output connector of the MU181500B and
the Ext. Clock Input connector of the MU183020A-x30 using a
coaxial cable.
7. Use a 130 cm coaxial cable (J1611A) to connect the Clock Output
connector of the MU183020A-x30 and the Ext Clock Input connector
of the MP1825B-x02.
8. Use an 80 cm coaxial cable (J1612A) to connect the Data Output
connector of the MU183020A-x30 and the Data Input connector of
the MP1825B-02.
9. Connect the Data Output and
Data
Output connectors of the
MP1825B-x02 and the Pick Off Tee (J1510A) (2 connections).
10. Connect the J1510A and the LFPS TX Output connector of the
USB3.1 Receiver Test Adapter with a 30 cm coaxial cable (J1624A)
as in Figure 3.4-1 (2 connections).
11. Connect the J1510A and Test Fixture with 80 cm coaxial cable
(J1612A) as in Figure 3.4-1 (2 connections).
12. Use a GND connector cable (J1627A) to connect the DUT GND and
MP1825B-x02 GND jacks.
13. Use a cable supplied with the USB3.1 Receiver Test Adapter to
connect the USB3.1 Receiver Test Adapter GND and MP1825B-x02
GND jacks.
14. Use a 100 cm coaxial cable (J1625A) to connect the AUX Output
connector of the MU183020A-x30 and the AUX Input connector of
the USB3.1 Receiver Test Adapter.
15. Use a 100 cm coaxial cable (J1625A) to connect the Gating Output
connector of the MU183020A-x30 and the Gating Input connector of
the USB3.1 Receiver Test Adapter.
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...