Chapter 5 Remote Control
5-74
:SENSe:JITTer:TABLe:INDex? <number>
Parameter
<number>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 50
Response
<string>=<STRING RESPONSE DATA>
<string>="<number>,<freq>,<mask>,<upperlimit><lowerlimit>"
<number>=<NR1 NUMERIC RESPONSE DATA>
1 to 50
Jitter Tolerance Table index No,1 to 50
<freq>=<NR1 NUMERIC RESPONSE DATA>
10 to 250000000
Modulation frequency registered in the Jitter
Tolerance Table
<mask>=<NR1 NUMERIC RESPONSE DATA>
0 to 2000
Mask value registered in the Jitter Tolerance
Table
<upperlimit>=<NR1 NUMERIC RESPONSE DATA>
0 to 2000
Upper Limit value registered in the Jitter
Tolerance Table
<lowerlimit>=<NR1 NUMERIC RESPONSE DATA>
0 to 2000
Lower Limit value registered in the Jitter
Tolerance Table
Function
Reads in values registered in the Jitter Tolerance Table by specifying the
Index.
Example
>:SENSe:JITTer:TABLe:IND? 1
< "1,150000000, 0.100, 0.200, 0.030"
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...