5.3 Command Description Method
5-11
5
R
em
ot
e C
on
trol
The following shows the description example of command.
■
Example of program command
:DISPlay:MEASure:CHANge <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
SETTing
Setting screen
RESult
Result screen
REPort
Report screen
Function
Switches the display screen for Tolerance/Sweep measurement.
Result cannot be specified for measurement not executed.
Example
Displays the Result screen for Tolerance measurement.
> :DISPlay:MEASure:CHANge RES
Program command
Parameter type name
Parameter type
Parameter contents
Command example
■
Example of query command
:SENSe:MEASure:SYSCond:SJSelect?
Response
<type>=<CHARACTER RESPONSE DATA>
OFF,SJ,SJ2
Function
Queries the sine wave jitter setting fixed and added for
Tolerance/Sweep measurement.
Example
> :SENSe:MEASure:SYSCond:SJSelect?
< SJ2
Command example, response example
Program command
Parameter type name
Parameter type
The < and > in the example indicate the response and the program
message respectively.
Notes:
●
Any commands for MX183000A are sequential commands.
●
If commands have restrictions, other settings may be affected.
For the setting items to be affected and conditions to be
restricted, refer to the
MX180000A Signal Quality Analyzer
Control Software Operation Manual and operation manual for
each module.
●
When the parameters of program command and query
command are same, the parameter of query command may be
omitted.
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...