5.10 Jitter Tolerance Setup Screen
5-85
5
R
em
ot
e C
on
trol
:SENSe:MEASure:BERCond:WTIMe <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 99
1 to 99 s/1s Step
Function
Sets the Waiting Time for the Tolerance measurement.
Example
To set the Waiting Time to 5 s:
> :SENSe:MEASure:BERCond:WTIMe 5
:SENSe:MEASure:BERCond:WTIMe?
Parameter
None
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
1 to 99
1 to 99 s
Function
Queries the Waiting Time of the Tolerance measurement.
Example
> :SENSe:MEASure:BERCond:WTIMe?
< 5
:SENSe:MEASure:BERCond:STIMe <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 99
1 to 99 s/1s step
Function
Set the Settling Time for the Tolerance measurement.
Example
To set the Settling Time to 5 s:
> :SENSe:MEASure:BERCond:STIMe 5
:SENSe:MEASure:BERCond:STIMe?
Parameter
None
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
1 to 99
1 to 99 s
Function
Queries the Settling Time of the Tolerance measurement.
Example
> :SENSe:MEASure:BERCond:STIMe?
< 5
:SENSe:MEASure:BERCond:GTIMe <time>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 86400
1 to 86400 s/1s step
Function
Set the Gating Time for the bit errors measurement.
Example
To set the Gating Time to 5 s:
> :SENSe:MEASure:BERCond:GTIMe 5
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...