Chapter 5 Remote Control
5-84
:DISPlay:RESult:ESTimate:ERATe <numeric1>,<numeric2>
Parameter
<numeric1>=<DECIMAL PROGRAM DATA>
1.0 to 9.9
XX:XXE-YY
<numeric2>=<DECIMAL PROGRAM DATA>
9 to 20
YY:XXE-YY
Function
Sets the BER for JTOL Estimation.
Example
To set the BER for JTOL Estimation to 5.5E–15:
> :DISPlay:RESult:ESTimate:ERATe 5.5,15
Compatibility
Incompatible with existing models.
:DISPlay:RESult:ESTimate:ERATe?
Response
<numeric1>,<numeric2>
<numeric1>=<DECIMAL RESPONSE DATA>
1.0 to 9.9
XX:XXE-YY
<numeric2>=<DECIMAL RESPONSE DATA>
9 to 20
YY:XXE-YY
Function
Queries the error rate setting set for BER for JTOL Estimation.
Example
> :DISPlay:RESult:ESTimate:ERATe?
< 5.5,15
Compatibility
Incompatible with existing models.
:SENSe:MEASure:BERCond:SEARch <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
BINary
Binary
DLINear
Downwards Linear
DLOG
Downwards Log
ULINear
Upwards Linear
ULOG
Upwards Log
BINLinear
Linear
Function
Sets the tolerance measurement method.
Example
To set the tolerance measurement method to Binary:
> :SENSe:MEASure:BERCond:SEARch BIN
:SENSe:MEASure:BERCond:SEARch?
Response
<type>=<CHARACTER RESPONSE DATA>
BIN, DLIN, DLOG, ULIN, ULOG, BINL
Function
Queries the tolerance measurement method.
Example
> :SENSe:MEASure:BERCond:SEARch?
< BIN
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...