5.9 USB Link Sequence Setup Screen (With Option 012 Installed)
5-59
5
R
em
ot
e C
on
trol
:LTRaining:SEQuence:TEST:PATTern <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
When Specification setting is GEN1:
0 to 6
CP0 to CP6
When Specification setting is GEN2:
9
CP9
Function
Selects the type of Compliance Pattern to be sent when test pattern is set
to Compliance.
Example
> :LTRaining:SEQuence:TEST:PATTern 0
:LTRaining:SEQuence:TEST:PATTern?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
When Specification setting is GEN1:
0 to 6
CP0 to CP6
When Specification setting is GEN2:
9
CP9
Function
Queries the test pattern to be sent.
Example
> :LTRaining:SEQuence:TEST:PATTern?
< 0
:LTRaining:SEQuence:TEST:PATTern:EADDition
Parameter
None
Function
Adds a single error to the sending test pattern.
Example
> :LTRaining:SEQuence:TEST:PATTern:EADDition
:LTRaining:SEQuence:INITialize [<spec>]
Parameter
<spec>=<CHARACTER PROGRAM DATA>
GEN1
GEN2
Note:
When <spec> is omitted, all parameters for the link training
sequences are initialized.
Function
Sets all parameters for the specified link training sequences to the initial
values.
Example
To initialize the parameter set by GEN2
> :LTRaining:SEQuence:INITialize GEN2
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...