Chapter 5 Remote Control
5-8
:SYSTem:MMEMory:RESult:STORe
"D:\test_folder\test",HTML
:SYSTem:MMEMory:RESult:STORe
"D:\test_folder\test",CSV
14. Close the application.
:SENSe:MEASure:JITTer:STOP
:SYSTem:MEASure:SELect NONE
This is an example of the procedure for jitter tolerance testing using a
PCIe device.
1. Start the PCIe Link Sequence application.
:SYSTem:MEASure:SELect PCI
2. Connect using steps 6 to 9 in the jitter tolerance test example
procedure.
3. Set the test target specification to Revision 3.
:DISPlay:MEASure:CHANge SEQuence
:LTRaining:SEQuence:SPECification REV3
4. Set the test pattern to Modified Compliance Pattern.
:SOURce:PATTern:TYPE COMPliance
:LTRaining:SEQuence:TEST:PATTern MCP
5. Send a sequence to loopback the DUT.
:LTRaining:SEQuence:STARt
6. During sequence transmission, the only remote control operations
possible are measurement stop and obtain measurement status.
:LTRaining:SEQuence:STATe?
:LTRaining:SEQuence:STOP
7. When the Option PL001 is installed, perform the jitter tolerance test
on the DUT in loopback status using steps 10 to 13 in the jitter
tolerance test example procedure.
8. Close the application.
:SENSe:MEASure:JITTer:STOP
:LTRaining:SEQuence:STOP
:SYSTem:MEASure:SELect NONE
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...