5.8 PCIe Link Sequence Setup Screen (With Option 011 Installed)
5-33
5
R
em
ot
e C
on
trol
:LTRaining:SEQuence:SPECification?
Response
<type>=<CHARACTER RESPONSE DATA>
REV1, REV2, REV3, REV4
Function
Queries the environment to loopback the DUT.
Example
> :LTRaining:SEQuence:SPECification?
< REV4
:LTRaining:SEQuence:LTHRough <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
CONFiguration
Configuration route
RECovery
Recovery route
Function
Select the LTSSM route to loopback the DUT.
Example
To set the state route to Configuration route:
> :LTRaining:SEQuence:LTHRough CONFiguration
:LTRaining:SEQuence:LTHRough?
Response
<type>=<CHARACTER RESPONSE DATA>
REC, CONF
Function
Queries the LTSSM route to loopback the DUT.
Example
> :LTRaining:SEQuence:LTHRough?
< CONF
:SOURce:PATTern:TYPE <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
COMPliance
Compliance pattern
PRBS
PRBS pattern
Function
Selects the test pattern to be sent after completing the link training
sequence transmission.
Example
To set the test pattern to Compliance Pattern:
>:SOURce:PATTern:TYPE COMPliance