Chapter 5 Remote Control
5-68
:SENSe:MEASure:BER:STOP
Parameter
None
Function
Starts the BER measurement.
Example
>:SENSe:MEASure:BER:STOP
:SENSe:MEASure:BER:STATe?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
0
Stop
1
Measurement in progress
Function
Queries the BER measurement status.
Example
>:SENSe:MEASure:BER:STATe?
< 1
:DISPlay:RESult:BER <boolean>
Parameter
<boolean>=<BOOLEAN PROGRAM DATA>
OFF or 0
Hides the BER measurement setting window.
ON or 1
Displays the BER measurement setting window.
Function
Displays or hides the BER measurement setting window.
Example
To display the BER measurement setting window
>:DISPlay:RESult:BER ON
:DISPlay:RESult:BER?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
0
Hides the BER measurement setting window.
1
Displays the BER measurement setting window.
Function
Queries the BER measurement setting window display status.
Example
>:DISPlay:RESult:BER?
< 1
:SENSe:MEASure:BER:ECTHreshold <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
0 to 1000
0 to 1000/1step
Function
Sets an evaluation threshold of the BER measurement.
When the number of bit errors exceeds the evaluation threshold, the BER
measurement is judged as Fail.
Example
To set the evaluation threshold of the BER measurement to 1:
>:SENS:MEAS:BER:ECTHreshold 1
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...