5.10 Jitter Tolerance Setup Screen
5-77
5
R
em
ot
e C
on
trol
:SENSe:MEASure:JITTer:STARt
Parameter
None
Function
Starts the tolerance measurement.
Example
>:SENSe:MEASure:JITTer:STARt
:SENSe:MEASure:JITTer:STOP
Parameter
None
Function
Stops the tolerance measurement.
Example
>:SENSe:MEASure:JITTer:STOP
:SENSe:MEASure:JITTer:STATe?
Response
<numeric>=<NR1 NUMERIC RESPONESE DATA>
1
Measurement being executed
0
Measurement stopped
Function
Queries the state of the Tolerance measurement.
Example
> :SENSe:MEASure:JITTer:STATe?
< 1
:CALCulate:RESult:STATus?
Response
<string>=<STRING RESPONSE DATA>
"---"
Before measurement start
"Setting system…"
Setup in progress
"Auto Search Started. "
Auto Search start
"Auto Search Completed. "
Auto Search complete
"Measurement Completed. "
Measurement complete
"Measurement: <Freq>Hz, <Currval>UIp-p"
Measurement result
Function
Queries the tolerance measurement status.
Example
> :CALCulate:RESult:STATus?
< "---"