:SYSTem:EQUipment:SETTing
<type>,<number>,<unit>,<slot>[,<data_if>]
Parameter
<type>=<CHARACTER PROGRAM DATA>
JITTer
Jitter Modulation Source
PPG
PPG
ED
ED
<number>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 3
MP1800A No.1 to 3
<slot>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 6
Slot 1 to 6
[<data_if>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4
Data 1 to 4
Note:
<data_if> can be set when the module installed in slot is bellow:
MU183020A, MU183021A, MU183040A/B, MU183041A/B
Data 1 is specified when omitted.
If <data_if> is set to other modules, the parameter error occurs.
With PCIe Link Sequence and USB Link Sequence, PPG <data_if>
can be set to "1" only.
Function
Selects the equipment to be used for the measurement.
Example
To assign MP1800A No.1, Unit 1, and Slot 4 to Jitter Modulation Source
:
> :SYSTem:EQUipment:SETTing JITTer,1,1,4
Assign MP1800A No. 1, Unit 1, Slot 4, Data 2 to the PPG.
> :SYSTem:EQUipment:SETTing PPG,1,1,4,2