3.3 PCIe Link Sequence Connection Procedure
3-11
3
C
on
nec
tin
g E
qu
ip
m
en
t
1
.
When MX183000A is installed on a control PC, connect the control
PC to MP1800A with an Ethernet cable. Or connect the MP1800A to
the MP1825B with a USB cable.
When MX183000A is installed on MP1800A, Ethernet cable
connection is not required.
MP1800A requires MP1800A-x02 LAN option.
2. Set up as follows using the Remote Control tab on Setup Utility.
Activate Interface:
Ethernet
Performance:
Enhanced
3. Mount MU181000A/B-x01 and MU181500B in MP1800A-x02, x32.
4. Mount MU183020A-x30 in Slot 3 of MP1800A-x02, x32.
5. Mount MU183040A/B in Slot 4 of MP1800A-x02, x32.
6. Connect the Clock Output connector of MU181000A/B-x01 to the Ext
Clock Input connector of MU181500B with a coaxial cable.
7. Use BNC-SMA cables (J1508A) to connect the Ext.I,Q connector of
the MU181000A/B-x01 and the I,Q Output connector of the
MU181500B. (2 connections)
8.
Connect the Jittered Clock Output connector of the MU181500B and
the Ext. Clock Input connector of the MU183020A-x30 using a
coaxial cable.
9.
Use coaxial cables to connect the Sub-rate Clock Output and XClock
Output of the MU181500B to the DUT REFCLK and INJ
connectors.
10.
Use a 130 cm coaxial cable (J1611A) to connect the Clock Output
connector of the MU183020A-x30 and the Ext Clock Input connector
of the MP1825B-x02.
11. Use an 80 cm coaxial cable (J1612A) to connect the Data Output
connector of the MU183020A-x30 and the Data Input connector of
the MP1825B-02.
12.
Connect MG3710A to 41KC-3, 41KC-6, 41KC-20, K241C and J1510A
as shown in Figure. 3.3-1. The MG3710A output is the RF Output
connector.
13. Set the MG3710A IP address, and connect to the PC.
14.
Connect the J1510A to the MP1825B-x02 Data Output and
Data
Output connector.
15.
Use an 80 cm coaxial cable (J1551A) to connect the J1510A and the
Compliance Base Board input (Rx In). (2 connections) Do not mount
the Addin Card at this stage.
16. Use a GND connector cable (J1627A) to connect the DUT GND and
MP1825B-x02 GND jacks.
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...