Chapter 5 Remote Control
5-18
:SYSTem:MEASure:SELect <item>
Parameter
<item>=<CHARACTER PROGRAM DATA>
NONE
Selector
TOLerance
Jitter Tolerance Test
PCI
PCIe Link Sequence
USB
USB Link Sequence
Function
Selects the application to be started.
Example
To select and start PCIe Link Sequence:
> :SYSTem:MEASure:SELect PCI
:SYSTem:MEASure:SELect <item>?
Response
<item>=<CHARACTER RESPONSE DATA>
NONE,TOL,PCI,USB
Function
Queries the application running.
Example
> :SYSTem:MEASure:SELect?
< PCI
:DISPlay:MEASure:CHANge <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
EQUipment
Setting window
SEQuence
Sequence screen
RUNTest
RunTest screen
GRAPh
Graph screen
REPort
Report window
Function
Switches between application display screens.
This command can be used after the application has been started using
the :SYSTem:MEASure:SELect command and connected to the MP1800A
using :SYSTem:EQUipment:CONNect.
Some screens may not be available depending on the particular
MX183000A license.
Example
To display the Tolerance Measurement Report screen:
> :DISPlay:MEASure:CHANge REPort