Chapter 5 Remote Control
5-50
:LTRaining:SEQuence:FSWing <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
12 to 63
12 to 63/1 Step
Function
Sets the TS Full Swing value.
Example
To set the TS Full Swing value to 30:
> :LTRaining:SEQuence:FSWing 30
:LTRaining:SEQuence:FSWing?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
12 to 63
12 to 63/1step
Function
Queries the TS Full Swing value.
Example
> :LTRaining:SEQuence:FSWing?
< 30
:LTRaining:SEQuence:LFRequency <numeric>
Parameter
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
12 to 63
12 to 63/1 Step
Function
Sets the TS Low Frequency value.
Example
To set the Low Frequency value to 30:
> :LTRaining:SEQuence:LFRequency 30
:LTRaining:SEQuence:LFRequency?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
12 to 63
12 to 63/1step
Function
Queries the TS Low Frequency value.
Example
> :LTRaining:SEQuence:LFRequency?
< 30
:LTRaining:SEQuence:DSCRamble <boolean>
Parameter
<boolean>=<BOOLEAN PROGRAM DATA>
OFF or 0
De-assert
ON or 1
Assert
Function
Sets the TS Disable scramble bit.
Example
To set Disable scramble to Asset:
> :LTRaining:SEQuence:DSCRamble 1
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...