Table A-7 Run Test Tab
Item
Specifications
Run Test/Stop Test
Detail
Check All
Uncheck
Starts or stops Jitter Tolerance Test.
Displays settings and results of Jitter Tolerance Table.
Select the Jitter Tolerance Table checkbox.
Deselect the Jitter Tolerance Table checkbox.
Jitter Tolerance Table
JTOL measurement point
Sets SJ modulation frequency to perform measurement, modulation
amount (UI) for pass/fail judgement, and range of modulation
amount to search.
Jitter frequency setup range
Sets Jitter Freq. [Hz], Mask[UI], Upper Limit [UI], Lower Limit
[UI], Upper Ratio, and Lower Ratio. Click the Add button to add
points. Click Delete > All Clear to delete points.
For setting range, refer to the sinusoidal jitter (SJ or SJ2)
modulation frequency setup range shown in Table 1.3-2 “Jitter
Modulation Performance” in the
MU181500B Jitter Modulation
Source Operation Manual.
Jitter amplitude setup range
Note that available jitter frequency and jitter amplitude for jitter
measurement depend on the clock frequency set by controller and
MU181500B.
Set All Limit
Resets the Upper Limit and Lower Limit values at the ratio set for
Mask.
Set the ratio to reset for Upper Ratio and Lower Ratio.
Upper Ratio
Lower Ratio
1.000 to 1000, 0.001 step
0.001 to 1.000, 0.001 step
Measurement Sequence
Mask file Save/Open
From higher Freq. side, From lower Freq. side
Saves and opens the mask file for JTOL test.
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...