:LTRaining:SEQuence:STARt
Parameter
None
Function
Starts transmitting a link training sequence for looping back DUT.
Once the transmission is complete, the pattern selected by Test Pattern
is sent.
Example
> :LTRaining:SEQuence:STARt
:LTRaining:SEQuence:STOP
Parameter
None
Function
Stops transmitting link training sequence and test pattern and sets to
Electrical Idle.
Example
> :LTRaining:SEQuence:STOP
:LTRaining:SEQuence:STATe?
Response
<numeric>=<NR1 NUMERIC RESPONSE DATA>
0
Stop
1
Sending
2
Sending test pattern
Function
Queries the link training sequence transmission status.
Example
> :LTRaining:SEQuence:STATe?
< 1
:LTRaining:SEQuence:SPECification <type>
Parameter
<type>=<CHARACTER PROGRAM DATA>
REV1
Revision1.0/1.1
REV2
Revision2.0
REV3
Revision3.0/3.1
REV4
Revision4.0
Function
Selects the environment (revision) to loopback the DUT.
Example
To set the environment to REV4(16.0 GT/s):
> :LTRaining:SEQuence:SPECification REV4
Note:
The clock frequency input to MU181500B must be changed by the
user when MU181000A/B is not installed.