Chapter 5 Remote Control
5-80
:INPut:CLOCk:SELection <sel>
Parameter
<sel>=<CHARACTER PROGRAM DATA>
RECovered
Recovered Clock
EXTernal
External Clock
Function
Sets the clock input type.
Example
To set the clock input type to the Recovered Clock:
> :INPut:CLOCk:SELection RECovered
Note:
This command is available when the application is started by PCIe
Link Sequence and the MU18304xB with Option 22 or 23 installed
is selected for Equipment.
Unavailable when the application is started by Jitter Tolerance
Test.
:INPut:CLOCk:SELection?
Response
<sel>=<CHARACTER RESPONSE DATA>
REC
Recovered Clock
EXT
External Clock
Function
Queries the clock input type.
Example
> :INPut:CLOCk:SELection?
< EXT
Note:
This command is available when the application is started by PCIe
Link Sequence and the MU18304xB with Option 22 or 23 installed
is selected for Equipment.
Unavailable when the application is started by Jitter Tolerance
Test.