R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
A.2 Error Message
A-6
85
Incorrect data.
Set span to (1.0 + alpha)*Tf or more.
The measurement condition is incorrect.
Change the measurement condition to
meet the following:
Frequency span > (1.0 + Rolloff factor)
×
Symbol rate
87
Root Nyquist filter is active.
Turn the filter off.
88
Separate-screen mode is active.
Change to Full-screen mode.
90
Not available.
Set XY anchor first.
95
Not available.
Turn off Artificial Analog mode.
96
Not available.
Turn on Artificial Analog mode.
100
Not available in High speed ADC mode.
105
Not available in Frequency Domain mode.
110
Not available in Continuous peak mode.
111
Not available in Continuous XdB Down mode.
115
Trigger source is not VIDEO or EXT.
Set trigger to VIDEO or EXT.
120
Not available.
Change the sweep time to less than 500sec.
125
Not available in Digital RBW mode.
Disabled while the digital RBW mode is
set.
126
Reached the limit of span in Digital RBW mode.
The value for SPAN is at the maximum
and cannot be increased while in the digital
RBW mode.
127
Not available.
Change span to less than the limit.
The current SPAN value is greater than the
limit value.
Reduce the value until it is below the limit
value which corresponds to the targeted
RBW.
300
Printer is not ready.
Check printer setup.
301
Printer cable problem.
Check printer cable.
302
Printer not responding.
304
SIO port is busy.
Serial I/O port is in operation.
Check to see if the item is properly set in
the RS232 dialog box under the Config
menu.
Code
Error message
Remarks