R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
3.3.12 MEAS Key (Measurement)
3-51
3.3.12 MEAS Key (Measurement)
This section describes the Measure menu displayed when the MEAS key is pressed.
Counter
Displays the Counter menu.
Resolution 1 kHz, 100 Hz, 10 Hz or 1 Hz
Allows you to set the resolution to 1 kHz, 100 Hz, 10 Hz or 1 Hz.
Counter ON/OFF
Toggles the frequency counter function on or off.
ON:
Measures the active marker frequency using the fre-
quency counter.
OFF:
Turns the frequency counter function off.
Peak List
Displays the Peak list menu.
Peak List Level
Lists the levels and frequencies in descending order of the peak
levels.
Peak List Freq
Lists the levels and frequencies in descending order of the peak
level frequencies.
List OFF
Turns off the peak list display function.
Noise/Hz
Displays the Noise/Hz menu.
dBm/Hz
Sets the vertical axis unit to dBm, and sets the marker readout sig-
nal level unit to dBm/Hz. In addition, the detector is automatically
set to Sample mode.
dB
µµµµ
V/
√√√√
Hz
Sets the vertical axis unit to dB
µ
V, and sets the marker readout
signal level unit to dB
µ
V /
√
Hz. In addition, the detector is auto-
matically set to Sample mode.
dBc/Hz
Sets the unit of Delta marker signal level to dBc/Hz and turns the
marker fixed function ON. In addition, the detector is automati-
cally set to Sample mode.
Fixed MKR Peak
Move the delta marker to the peak currently displayed (on the
trace) in order to make it fixed in this position.
Noise/Hz OFF
Turns off the noise measurement mode and returns to the Measure
menu.
XdB Down
X dB Down menu is displayed to allow you to set the attenuation.
X dB Down
Moves Normal and Delta markers to an intersection point on the
trace X dB down from the present location.