R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
A.2 Error Message
A-7
305
Input buffer overflow.
The input buffer overflowed.
Send the data again to the input buffer.
400
Input ATT Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
401
IF Step AMP Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
402
Log Linearity Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
403
Total Gain Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
404
RBW Switching Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
405
Amplitude MAG Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
406
Insufficient Cal data.
Execute CAL ALL.
Cannot be corrected because of insufficient
correction conditions.
Perform CAL ALL.
409
Normal ADC Cal failed.
The calibration failed.
Run the user self test to check the problem
again.
600
Illegal parameter(s).
601
Illegal file or device name.
602
Incompatible firmware version.
Data cannot be used with this analyzer.
603
Cannot be formatted.
The device cannot be formatted.
Try to format it using personal computer
and so on.
604
Cannot rename this file.
The file name on a RAM disk cannot be
changed.
605
Corrupt file data.
The saved data cannot be used since it is
corrupt.
607
Specified device does not exist.
The specified device name cannot be
found.
Use the correct device name.
608
No media present.
The device is not ready for operation.
Insert a floppy disk or the memory card.
609
Directory not found.
Code
Error message
Remarks