R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.5 Entering Level Correction Data
2-30
Figure 2-32 Setting Measurement Conditions
Entering the correction data
The correction table is composed of frequency and level columns, and is used to set a
maximum of 50 sets of data. The interpolation method is applicable between correc-
tion data.
6.
Press LEVEL, Correction Factor and Correction Edit.
The Correction Edit menu and the correction table are displayed and you are
allowed to enter frequency data (See Figure 2-33).
Figure 2-33 Displaying the Correction Table
7.
Press 6, 6, 0 and MHz.
A frequency of 660 MHz is displayed in the first frequency item, and the cursor
moves to the level item (See Figure 2-34).