R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
3.3.16 POWER Key (Power Measurement)
3-60
√√√√
Nyquist Filter
Toggles the Nyquist filter function on or off.
ON:
Turns the Nyquist filter function on.
OFF:
Turns the Nyquist filter function off.
Parameter Setup
Displays the ACP Setup menu.
Default
Reads the frequency span, RBW, VBW, the sweep time and the
status of the detector which have previously been saved using
"Define
→
Default."
Manual
Sets the channel space and channel bandwidth.
Define
→
→
→
→
Default Registers the frequency span, RBW, VBW, the sweep time and
the status of the detector, which have previously been set manual-
ly, as the preset values.
Screen FULL/SEPA/CARRIER
Toggles the screen display between the full and separate screen
modes.
FULL:
The entire screen is used.
Calculates the power in reference to the band of the en-
tire screen.
SEPA:
The separate screen is used.
CARRIER:
The entire screen mode is used. Calculates the power in
reference to the power of band specified by the carrier
band.
Average Times ON/OFF
Toggles the average function on or off.
ON:
Sets the number of times averagings are performed, and
the measures the average adjcent channel leakage pow-
er.
OFF:
Measures the ACP on a sweep basis.
Graph
Displays the ACP Graph menu.
Graph ON/OFF
Toggles the graph display on or off.
ON:
Displays the leakage power graph as Trace B and puts
Delta marker in the center of the screen.
The B memory is used to display the ACP graph.
OFF:
Turns off the graph display.