R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.4 Display Line and Measuring Window
2-27
Figure 2-29 The Display Line
8.
Align the Display line to a peak on the right.
9.
Press SRCH.
A marker is displayed on the trace peak.
10. Press MKR, Reference Object and Display Line.
The values shown by the marker are the values relative to the Display line.
Figure 2-30 Measuring the Values Relative to the Display Line
Removing the Display line
11. Press FORMAT, Display Line ON/OFF(ON) and Display Line ON/OFF(OFF).
The Display line and the values relative are removed.