R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.4 Display Line and Measuring Window
2-26
2.2.4
Display Line and Measuring Window
This section describes the display line used to compare the levels between traces and the measuring win-
dow used to take measurements within a limited area.
Power on
1.
Turn the analyzer power on.
Initialization
This resets the current settings to the factory defaults or user-defined presets.
2.
Press SHIFT and CONFIG(PRESET).
This sets the analyzer to its presets values.
Connecting the calibration signal
Connect the calibration signal used in the measurement.
3.
Attach the N-BNC adapter to the INPUT connector on the front panel.
4.
Connect the Input cable from the CAL OUT connector to the INPUT connector.
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
5.
Press FREQ, 3, 0 and MHz.
A center frequency of 30 MHz is set.
6.
Press SPAN, 8, 0 and MHz.
A frequency span of 80 MHz is set.
Turning on the display line
7.
Press FORMAT and Display Line ON/OFF(ON).
The display line is displayed.