R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
5.2.9 GPIB Command Codes
5-44
Table 5-22 Miscellaneous
Function
Listener
Code
Talker Request
Panel
Key
Remarks
Code
Output Format
Miscellaneous Outputting error number
–––
ERRNO?
Integer
Refer to the error
number found in
the Error Message
List.
Local
LC
–––
–––
LCL
Reading GPIB address
–––
AD?
Integer (0 to 30)
Specification of the delimiter CR LF <EOL> DL0
–––
–––
LF
DL1
–––
–––
<EOI>
DL2
–––
–––
CR LF
DL3
–––
–––
LF <EOI>
DL4
–––
–––
Service request interruption ON
S0
–––
–––
OFF
S1
–––
–––
Status clear
S2
–––
–––
Service request mask
RQS *
RQS?
Decimal number correspond-
ing to the SRQ bit
Outputting ID of the instrument
–––
*IDN?
Manufacturer name (charac-
ter string), instrument type
(character string), Serial
number (character string) and
revision (character string)
Initializing the instrument
*RST
–––
–––
Clearing the queues related to the status byte *CLS
–––
–––
Accessing the standard event enable register *ESE
*ESE?
Decimal number correspond-
ing to the register bits
Reading or clearing the standard event
enable register
–––
*ESR?
Decimal number correspond-
ing to the register bits
Accessing the service request enable register *SRE
*SRE?
Decimal number correspond-
ing to the register bits
Reading the status byte and MSS bit
–––
*STB?
Decimal number correspond-
ing to the status byte
Accessing the operation status enable regis-
ter
OPR
OPR?
Decimal number correspond-
ing to the register bits
Reading or clearing the operation status reg-
ister
–––
OPREVT? Decimal number correspond-
ing to the register bits
Reading of the self test result
–––
*TST?
0: Pass
1: Power Up & CPU Block
2: Synthe Block
4: RF Block
8: IF Output
16: Log/AD Block
32: IF BLOCK
For the item
which resulted in
an error, the
return value is the
value obtained by
ORing the related
bits.
PIO data output
PIOOUT *
–––
–––
Outputs data with
respect to 0 to 255
to the PARAL-
LEL port pins 2
thru 9 as a set of
eight-bit data.