![ADVANTEST R3267 series Operation Manual Download Page 139](http://html1.mh-extra.com/html/advantest/r3267-series/r3267-series_operation-manual_2858766139.webp)
R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.3.8 Measuring Burst signals in the Time Domain
2-102
7.
Press LEVEL, 5 and GHz(+dBm).
The reference level is set to +5 dBm.
8.
Press COUPLE, RBW AUTO/MNL(MNL), 3 and MHz.
An RBW of 3 MHz is set.
The burst signal used with TDMA can be checked.
Figure 2-112 Burst signal in the frequency domain
9.
Press SPAN and Zero Span.
The frequency span is set to zero span.
10. Press SWP, Sweep Time AUTO/MNL(MNL), 5 and kHz(ms).
A sweep time of 5 msec is set.
Figure 2-113 Burst Signal in the Time Domain
Setting the video trigger
11. Press Trigger Setup.
The Trigger Setup dialog box is displayed.