R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
3.3.3 B Key (Trace B)
3-25
3.3.3
B Key (Trace B)
This section describes the Trace B (1) menu used for the trace function displayed when the B key is
pressed.
Write B
This mode displays trace data from memory B, which is updated
for each sweep.
View B
This mode displays trace data previously saved in memory B.
Blank B
This mode erases the trace data currently stored in memory B.
Average B
Allows you to set the number of times the sweep is performed for
averaging. Once sweeping has begun, the result for each sweep
(which averaged with the previous settings) is displayed until the
set count is reached.
Max Hold B
Allows you to set the number of times the sweep Max Hold is per-
formed. Once sweeping has begun, the maximum result for each
sweep is kept and displayed until the set count is reached.
Trace B Detector
Displays the Detector B menu.
Normal
Sets the normal detector mode which automatically detects posi-
tive or negative peaks for each trace point.
Positive
Sets the positive peak detector mode.
Negative
Sets the negative peak detector mode.
Sample
Sets the sample detector mode.
DET Select AUTO/MNL
Toggles the detector mode between AUTO (automatic) or MNL
(manual) settings.
AUTO:
Automatically sets to one of the following detector
modes under which the most appropriate detector is ob-
tained.
The detector mode will not change if the Trace mode is
in Write mode.
MNL:
Sets the detector mode to manual mode.
more 1/2
Displays the Trace B (2) menu.
Trace mode
Detector mode
Average A
Sample
Max Hold A
Positive
Min Hold A
Negative
Power Average A
Sample