R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.3.4 Measuring CDMA Channel Power
2-83
15. Press Channel Width, 1, ., 2, 2, 8 and MHz.
The width of the measuring window is set to 1.228 MHz.
16. Press Average Times, 1, 0 and Hz.
An averaging count of 10 is set.
The channel power and the averaging count are displayed in the result area (See
Figure 2-92).
Figure 2-92 Result of Channel Power Measurement