R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.2 Displaying Spectrums and Operating the Markers
2-19
Figure 2-18 Factory Defaults
Connecting calibration signal
Connect the calibration signal used in the measurement.
7.
Attach the N-BNC adapter to the INPUT connector on the front panel.
8.
Connect the Input cable from the CAL OUT connector to the INPUT connector.
Figure 2-19 Calibration Output
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
9.
Press FREQ.
The current center frequency is displayed in the active area, and the Freq menu
used to select the frequency type appears on the right.