![ADVANTEST R3267 series Operation Manual Download Page 103](http://html1.mh-extra.com/html/advantest/r3267-series/r3267-series_operation-manual_2858766103.webp)
R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.14 Pass/Fail Judgments Using the Limit line Function
2-66
Setting the limit line
Each limit line uses the data in the table.
Table 2-3 Setting Limit Line 1
Table 2-4 Setting Limit Line 2
8.
Press FORMAT, Limit Line and Limit Line Edit.
The Edit menu and editor used for Limit Line 1 are displayed.
Figure 2-73 Editing the Limit Line 1
9.
Press 2, 0 and MHz.
20 MHz is set in the first frequency entry, and the cursor moves to the level entry.
10. Press 7, 0 and MHz(-dBm).
-70 dBm is set in the first level entry, and the cursor moves to the second row.
Frequency
Level
1
20 MHz
-70 dBm
2
26 MHz
-70 dBm
3
28 MHz
-5 dBm
4
32 MHz
-5 dBm
5
34 MHz
-70 dBm
6
40 MHz
-70 dBm
Frequency
Level
1
20 MHz
-95 dBm
2
28 MHz
-95 dBm
3
29 MHz
-60 dBm
4
31 MHz
-60 dBm
5
32 MHz
-95 dBm
6
40 MHz
-95 dBm