R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.3.6 Measuring Adjacent Channel Leakage Power (ACP)
2-91
23. Set
√√√√
Nyquist Filter ON/OFF(ON).
Allows you to set parameters and displays the data enter.
Figure 2-98 Root Nyquist Filter dialog box
24. Press
√√√√
Nyquist Filter Setup.
This closes the dialog box used for setting Root Nyquist Filter parameters.
Performing ACP
25. One marker is displayed in each of the upper and lower adjacent channels each
time a sweep is performed, and the lower adjacent channel leakage power as well
as the upper adjacent channel leakage power is displayed.
When you press SINGLE, only one measurement is taken.
Figure 2-99 ACP Measurement Display in Full Screen Mode