R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.5 Entering Level Correction Data
2-31
Figure 2-34 Entering Frequency Data
8.
Press 2, 0 and MHz(-dBm).
A level of -20 dBm is displayed in the level item and the cursor moves to the fre-
quency item on the second line.
9.
Enter the correction data one by one according to Table 2-1.
10. Press RETURN.
The correction table is removed.
Reflecting the level correction data
11. Press Correction ON/OFF(ON).
The trace, whose noise level was corrected using the data previously entered, is
displayed.
Figure 2-35 Showing a Trace Whose Level Is Corrected