Operation
Instrument Controls Menu
3-50
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Concentrations
Other Measurements
Analog Inputs
LO NO
2
dual mode)
Chamber Pressure
Analog Input 5
LO NO
x
(dual mode)
Flow
Analog Input 6
HI NO (dual mode)
PMT Volts
Analog Input 7
HI NO
2
(dual mode)
Ozonator Flow
Analog Input 8
HI NO
x
dual mode)
Range (NO
x
) (auto range mode)
Prereactor (single/auto range mode)
LO Prereactor (dual mode)
HI Prereactor (dual mode)
The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O
expansion board option is installed. Configuration includes entering
descriptor, units, decimal places, choice of 1-10 points in the table, and
corresponding number of points selected.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config
.
ANALOG INPUT CONFIG:
>CHANNEL 1 IN1
CHANNEL 2 IN2
CHANNEL 3 IN3
CHANNEL 4 IN4
CHANNEL 5 IN5
CHANNEL 6 IN6
CHANNEL 7 IN7
ª
RANGE AVG DIAGS ALARM
ANALOG INPUT 01 CONFIG:
>DESCRIPTOR IN1
UNITS V
DECIMAL PLACES 2
TABLE POINTS 2
POINT 1
POINT 2
RANGE AVG DIAGS ALARM
The Descriptor screen allows the user to enter the descriptor for the
selected analog input channel. The descriptor is used in datalogging and
Analog Input
Configuration
Descriptor