Servicing
Pressure Transducer Assembly Replacement
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
7-35
3.
At the Analog Input Cal screen, press
to scroll to the channel
selected in Step 1, and press
.
4.
Press
to scroll to Calibrate Full Scale, and press
.
The screen displays the current input voltage for the selected channel.
5.
Use
and
to enter the source voltage, and press
to calibrate the input voltage for the selected channel to the
source voltage.
6.
Press
>
to return to the input channels display and repeat
Steps 3-5 to calibrate other input channels to the source voltage as
necessary.
Use the following procedure to replace the pressure transducer assembly
(
Figure 7–17
).
Equipment Required:
Pressure transducer assembly
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
2.
Disconnect plumbing from the pressure transducer assembly. Note the
plumbing connections to facilitate reconnection.
3.
Disconnect the pressure transducer cable.
Pressure Transducer
Assembly Replacement