Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-35
The Memory Allocation Percent screen is used to select the percentage of
each record type for both short records and long records. Percentages
between 0 and 100% are available in increments of 10. Changing this value
results in log erasure for both short records and long records.
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Memory Allocation %
.
SET PERCENT LRECS:
CURRENTLY: 50%
SET TO: 60% ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Data Treatment screen is used to select the data type for the selected
record type: whether the data should be averaged over the interval, the
minimum or maximum measured during the interval, or the current value
(last value measured). Data treatment doesn’t apply to all data, just to the
concentration measurement. All other data points log the current value at
the end of the interval.
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Data Treatment
.
SET LREC DATA TYPE:
CURRENTLY: AVG
SET TO: CUR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Communication Settings menu is used with communications control
and configuration.
●
In the Main Menu, choose Instrument Controls >
Communication
Settings
.
Memory Allocation
Percent
Data Treatment
Communication
Settings