C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
B-13
Figure B–1.
Flags
alarm chamber temp min
alarm chamber temp max
These commands report the chamber temperature alarm minimum and
maximum value current settings. The example below reports that the
chamber temperature alarm minimum value is 47.0 °C.
Send:
alarm chamber temp min
Receive:
alarm chamber temp min 47.0 deg C
set alarm chamber temp min
value
set alarm chamber temp max
value
These commands set the chamber temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing chamber temperature alarm limits in degrees C. The example
below sets the chamber temperature alarm maximum value to 50.0 °C.
Send:
set alarm chamber temp max 50.0
Receive:
set alarm chamber temp max 50.0 ok
Alarms