Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-43
I/O CONFIGURATION:
>OUTPUT RELAY SETTINGS
DIGITAL INPUT SETTINGS
ANALOG OUTPUT CONFIG
ANALOG INPUT CONFIG
RANGE AVG DIAGS ALARM
The Output Relay Settings menu displays a list of the 10 digital output
relays available, and allows the user to select the logic state or instrument
parameter for the relay selected.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings
.
OUTPUT RELAY SETTINGS:
>1 NOP GEN ALARM
2 NOP NONE
3 NOP UNITS
4 NOP CONC ALARM
5 NOP NONE
6 NOP NONE
7 NOP NO MODE
ª
RANGE AVG DIAGS ALARM
The Logic State screen is used to change the I/O relay to either normally
open or normally closed.
●
Press
to toggle and set the logic state open or closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
The Instrument State submenu allows the user to select the instrument
state that is tied to the selected relay output. A submenu lists signal types of
either alarm or non-alarm to choose from.
Output Relay Settings
Logic State
Instrument State