Operation
Calibration Menu
3-20
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
CALIBRATION PRESSURE:
PRESSURE: 232.5
SET CAL PRES TO: 220.0
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
●
In the Main Menu, choose Service > Calibration Factors >
Reset User
Cal Defaults
.
The Calibration menu is used to calibrate the analyzer, including zero
backgrounds, the NO, NO
2
, and NO
x
coefficients, and the calibration
pressure. The screens below show the calibration menu in single mode and
dual/auto range modes. The zero/span check is visible only if the zero/span
option is installed in single range mode.
The calibration procedure is the same in dual, auto, or single range,
however, there are two sets of gas coefficients in dual or auto range (i.e. low
and high coefficients). This enables each range to be calibrated separately.
When calibrating the instrument in dual or auto range, be sure to use a low
span gas to calibrate the low range and a high span gas to calibrate the high
range.
Whether dual or singe range, the first step in calibrating the instrument is
to assign the calibration pressure. The calibration pressure is used to
account for slight pressure fluctuation brought on by changing weather.
The calibration pressure is set equal to the reactor pressure at the time of
Reset User Calibration
Default
Calibration
Menu
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM