Servicing
Flow Transducer Calibration
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
7-47
WARNING
The service procedures in this manual are restricted to qualified
service representatives.
▲
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Remove the cover.
2.
Disconnect the pump cable from AC PUMP connector on the
measurement interface board.
3.
From the Main Menu, press
to scroll to Service > press
>
to scroll to
Flow Calibration
> and press
.
The Flow Sensor Cal menu appears.
Note
If Service Mode is not displayed, refer to “
Accessing the Service
Mode
” on page
7-4
, then return to the beginning of this step.
▲
4.
At the Flow Sensor Cal menu, press
to select
Zero
.
The Calibrate Flow Zero screen appears.
5.
Wait at least 10 seconds for the zero reading to stabilize, then press
to save the zero flow value.
6.
Reconnect the pump cable to the AC PUMP connector on the
measurement interface board.
7.
Connect a calibrated flow sensor at the SAMPLE bulkhead on the rear
panel.
8.
Press
to return to the Flow Sensor Cal menu.