Operation
Alarms Menu
3-68
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
The Flow screen displays the current sample flow reading and sets the
minimum and maximum alarm limits. Acceptable alarm limits range from
0 to 2 LPM. If the sample flow reading goes beyond either the minimum
or maximum limit, an alarm is activated. The word “ALARM” appears in
the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Flow
.
FLOW:
ACTUAL 0.531 LPM
>MIN 0.750 LPM
MAX 1.500 LPM
RANGE AVG DIAGS ALARM
The Minimum Flow alarm limit screen is used to change the minimum
sample flow alarm limit. The minimum and maximum flow screens
function the same way.
●
In the Main Menu, choose Alarms > Flow >
Min
or
Max
.
FLOW:
ACTUAL MIN: 0.750 LPM
SET MIN TO: 0.760 LPM ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Ozonator Flow screen (read only) is used to display the ozonator flow
readings. If the ozonator flow reading is 0.050 LPM (50 cc) or below, an
alarm is activated, and an alarm condition screen appears as “LOW”. If the
ozonator flow is above 0.050, the no alarm condition screen is displayed,
indicating that the flow is acceptable. Inadequate ozonator flow will cause
the ozonator to overheat, resulting in permanent damage to the ozonator.
●
In the Main Menu, choose Alarms >
Ozonator Flow
.
Flow
Min and Max Flow Limits
Ozonator Flow