Servicing
Input Board Replacement
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
7-27
2.
Remove the ozonator assembly as described in “Ozonator Assembly
Replacement.”
3.
Disconnect the plug connecting the ozonator transformer to the
measurement interface board (OZONATOR connector).
4.
Remove the four screws holding the ozonator transformer to the
ozonator bracket and remove the ozonator transformer.
5.
To install the ozonator transformer, follow the previous steps in reverse.
6.
Re-install the measurement bench. Refer to “
Removing the
Measurement Bench and Lowering the Partition Panel
” in this chapter.
Use the following procedure to replace the input board (
Figure 7–14
).
Equipment Required:
Input board
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Refer to “
Removing the Measurement Bench and Lowering the
Partition Panel
” in this chapter to lower the partition panel, then
proceed to the next step below.
2.
Disconnect the coaxial cable with BNC connector and the ribbon
cable.
3.
Loosen the two screws holding the assembly bracket to the floor plate,
move the assembly towards the rear, and lift the assembly off the
screws.
4.
Loosen the two screws holding the input box to the assembly bracket
and lift the input box off the screws.
Input Board
Replacement