Servicing
Input Board Calibration
7-28
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Figure 7–14.
Replacing the Input Board
5.
Remove the four screws holding the input board to the input box and
remove the input board.
6.
Install the input board by following the previous steps in reverse.
7.
Re-install the measurement bench. Refer to “
Removing the
Measurement Bench and Lowering the Partition Panel
” in this chapter.
8.
Perform an input board calibration. See the “Input Board Calibration”
procedure that follows.
After replacing the input board, use the following procedure to calibrate the
input board.
WARNING
The service procedures in this manual are restricted to qualified
service representatives.
▲
1.
From the Main Menu, press
to scroll to Service > press
>
to scroll to
Input Board Calibration
> and press
.
The Input Board Calibration screen appears.
Input Board
Calibration
Assembly Bracket
Input Board
Input Board Screws (4)
Input Box Screws (2)
Assembly Bracket Screws (2)
Input Box