Solenoid Valve Replacement
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
7-23
Figure 7–11.
NO
2
-to-NO Molybdenum Converter Assembly
6.
Remove the six screws holding the top housing assembly to the bottom
half.
7.
Remove the converter cartridge/heater assembly from the bottom
housing assembly.
8.
Loosen the heater clamp, pry heater apart no wider than necessary and
remove the converter cartridge noting the proper orientation of heater
wires and thermocouple probe.
9.
To replace converter, follow previous steps in reverse.
Note
Be sure to
wrap the O
3
converter tube snugly around the heater.
Use the following procedure to replace a solenoid valve (
Figure 7–12
).
Equipment Required:
Solenoid valve
Wrench, 5/16-inch
Philips screwdriver
Solenoid Valve
Replacement
Top Housing Assembly
Heater Assembly
Thermocouple Leads
and Heater Connector
Bottom Housing Assembly
Captive Screws (4)
O
3
Converter
Moly Converter Assembly