Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
D-4
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
The <address> is optional, which means it can be left out completely. The
<address> if present must match the Instrument Address. Additional space
can be present after the <address>.
If the received command does not satisfy the above formats or if the
<address> does not match the Instrument Address the command is ignored.
This is a sample command to switch the instrument to zero mode,
instrument id 5:
<STX>ST005<SP>N<CR>
This command DA initiates a data transfer from the instrument. The
instrument responds with measurement data, which depends on the range
mode and is listed in “
Measurements reported in response to DA
command
” below.
The command structure for a data query command is as follows:
<STX>DA<address><ETX><BCC>
The <address> is optional, which means it can be left out completely. The
<address> if present must match the Instrument Address. Additional space
can be present after the <address>.
If the <address> is left out then no space is allowed in the query string.
A command with no address is also a valid command.
The following are the different acceptable formats of the DA command
with Instrument Address 5:
<STX>DA<CR>
<STX>DA005<CR>
<STX>DA<SP><SP>5<ETX><BCC>
<STX>DA<ETX><BCC>
The data query string is valid and will be answered with data transmission
only if the command starts with <STX> which is followed by the characters
DA, and the <address> (if present) matches the Instrument Address, and
the command is terminated with either <CR> with no checksum or <ETX>
followed by the correct checksum <BCC>.
Data Sampling/Data
Query Command
(DA)