Operation
Service Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-75
SET CONVERTER TEMP:
CURRENTLY: 325
o
C
SET TO: 325
o
C
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure Calibration submenu is used to calibrate the pressure sensor
to zero, span, or restore factory default values. The pressure calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see “
Service Mode
” earlier in this chapter.
The pressure sensor’s zero counts and span slope are displayed on the
menu.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Pressure Calibration
.
PRESSURE SENSOR CAL:
>ZERO 72
SPAN 1.1416
SET DEFAULTS
RANGE AVG DIAGS ALARM
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note
A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
▲
●
In the Main Menu, choose Service > Pressure Calibration >
Zero
.
Pressure Calibration
Calibrate Pressure Zero