Operation
Service Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-81
The Temperature calibration screen allows the user to view and set the
ambient temperature sensor calibration. The temperature calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Temperature Calibration
.
CALIBRATE AMBIENT TEMP:
CURRENTLY: 32.3
o
C
SET TO: 032.
5
o
C ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Analog Output Calibration menu is a selection of 6 voltage channels
and 6 current channels (if I/O expansion board option is installed) to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration is visible only when the instrument is in
service mode. For more information on the service mode, see “
Service
Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Analog Out Calibration
.
Temperature
Calibration
Analog Output
Calibration
INPUT BOARD CALIBRATION:
TO CALIBRATE
** WARNING **
CONCENTRATION CALCULATION
x
xx
IS HALTED INSIDE
XX XX
THIS SCREEN!
XX
RANGE AVG DIAGS ALARM
INPUT GAIN TEST:
GAIN = 1
TEST = OFF
FREQ = 5000
§¨
CHG GAIN
©©
TEST MODE
RANGE AVG DIAGS ALARM