Operation
Range Menu
3-14
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Table 3–5
lists the available operating ranges.
●
In the Main Menu, choose Range >
NO, NO2, or NOx Range
.
RANGE:
CURRENTLY: 50
SET TO: 100 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Table 3–5.
Available Operating Ranges
ppb
μgm
3
5 10
10 20
20 50
50 100
100 200
200 500
C1 C1
C2 C2
C3 C3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “
Set Custom Ranges
” below.
The Custom Ranges Menu lists three custom ranges: C1, C2, and C3.
Custom ranges are user-defined ranges. In the ppb mode, any value
between 5 and 200 ppb can be specified as a range. In the μg/m
3
mode, any
value between 10 and 500 μg/m
3
can be specified as a range.
●
In the Main Menu, choose Range >
Set Custom Ranges
.
Set Custom Ranges